TESTING
36
®
TESTING
37
www.bayerdiabetes.com/us
Need Help? Call toll free: 1.800.348.8100 (available 24 hours a day, 7 days a week)
Accepted
10:30 AM | 13/11
121
mg
dL
Control Test
Control Solution Testing
Control Solution Testing
The result will automatically be marked as a control
test and stored in the meter memory. Control results
will not be included in your blood sugar average or
sent to your pump. To turn your meter off, hold down
the Menu button for 2 seconds.
If your control test result falls out of the specified
range, call Bayer Diabetes Care Customer Service
at 1-800-348-8100 (available 24 hours a day, 7 days
a week). Do not use the meter for blood glucose
testing until you resolve this issue.
3. Prepare to run a control test by inserting a test strip
into the meter.
4. Gently rock the control bottle before opening to
ensure it is mixed well.
5. Squeeze a small drop of control solution onto a
clean non-absorbent surface, such as a piece of
wax paper. Do not apply control solution to
your fingertip or to the test strip directly from
the bottle.
6. Replace the cap on
the control solution
bottle.
7. Immediately touch the
tip of the test strip to
the drop of control
solution.
8. Hold it in the drop
until the meter beeps.
9. Compare your control test result with
the control range. The Level 1 and
Level 2 control range values can be
found on the test strip bottle or carton.
84359751_CntrNextLINK_UG_EN_FpBp_v4.indd 36-3784359751_CntrNextLINK_UG_EN_FpBp_v4.indd 36-37 3/29/12 12:19 PM3/29/12 12:19 PM